A high-performing model still has to be validated before it can serve as an imaging biomarker. Sponsors need to establish where a quantitative PET result comes from, how reliably it travels across sites and what clinical decision it can support.
The most persuasive artificial intelligence demonstration is often visual: a model outlines a lesion in seconds, reconstructs an image from fewer counts or produces a crisp numerical output. The most important clinical-trial question comes later: if the same participant were assessed at another centre, on another qualified scanner or after an authorised software update, would the endpoint tell the same scientific story?
For pharmaceutical and biotechnology sponsors, AI-assisted quantitative PET is a measurement-system problem before it is a product-feature story. Imaging biomarkers may help characterise treatment effects, identify patient subgroups or support exploratory drug-development decisions. They may also introduce bias if acquisition, processing, segmentation or statistical interpretation differ between datasets. A fast algorithm still needs a defensible measurement strategy.
Consider three hypothetical development uses for a PET-derived metric. In the first, an automated lesion measurement helps an imaging team prioritise QC review; the immediate consequence of error may be a manual correction. In the second, a metric determines whether a patient is eligible to join a study; an error may change the study population. In the third, a quantitative metric contributes to a pivotal efficacy endpoint; error may undermine the primary estimate of treatment effect. These uses cannot share the same evidentiary threshold merely because the same software produces the number.
Regulators use the term context of use to describe the specific purpose and circumstances in which a biomarker or drug development tool is intended to be applied [1]. A sponsor should be able to write one defensible sentence stating the imaging variable, population, tracer, acquisition condition, intended decision and whether the output is supportive, exploratory or endpoint-defining.
Example context-of-use statement (illustrative only): “An automated PET-derived lesion-volume measurement, obtained from the specified tracer and qualified scanner/reconstruction settings, will be evaluated as an exploratory pharmacodynamic biomarker in adults with [defined disease] receiving [investigational treatment].” The bracketed elements require product-specific decisions. Such a statement does not imply FDA biomarker qualification or acceptance as a surrogate endpoint.
PET signal begins with tracer biology, but the observed measurement is shaped by patient preparation, administered activity, uptake interval, acquisition, attenuation and scatter correction, reconstruction, calibration, motion, anatomical localisation and the chosen analysis method. Static uptake metrics, volumetric measures and dynamic kinetic parameters have distinct error sources. The more complex the quantity, the more important it becomes to define its full provenance.
For example, two scans with identical biological uptake may produce meaningfully different apparent lesion metrics if the injection-to-acquisition (uptake) intervals, image resolution or segmentation approaches differ. Small lesions can be particularly vulnerable to partial-volume effects and segmentation instability. When a trial compares serial images, technical drift can masquerade as biological response or obscure a real change.
A modern imaging core lab should consequently treat scanner qualification, image QC, reconstruction specifications and software governance as part of endpoint development. The goal is evidence that measurements are comparable across sites and scanners.
Cardiac PET offers a useful precedent: tools that measure myocardial blood flow have faced the question of whether different software gives the same number for more than a decade. Professor Piotr J. Słomka has contributed to quantitative nuclear imaging and AI-based cardiac imaging research. The studies cited here are a few examples of cardiac PET literature, chosen because they address the question at the center of this article: whether different software gives the same number. His 2012 work with collaborators compared clinical tools for measuring regional myocardial blood flow with 13N-ammonia PET/CT [2]. A later study co-authored with Professor Juhani Knuuti assessed software reproducibility of myocardial blood-flow and flow-reserve quantification [3]. The first found good agreement between the packages; the second found that, in patients with ischaemia, stress flow and flow reserve differed between packages beyond accepted limits of agreement. Agreement between quantitative tools therefore has to be tested in each setting where they are used.
The same question applies to AI. In two 2024 reviews, Robert J. H. Miller and Professor Słomka described AI developments in nuclear cardiology, including opportunities to assist image acquisition, processing and interpretation [4,5]. Those methods were developed and tested in cardiac imaging, but the lesson applies to all PET imaging: an AI tool, like any quantitative tool, has to be tested in the population and setting where it will be used.
At Axcellant, this means a quantitative tool is treated as qualified only for the trial, tracer and scanner mix in which its performance has been checked. Published results from other settings help decide what to test; trial-specific data then show whether the tool is fit for use. The four questions below set out how.
Begin with engineering validity: DICOM compatibility, orientation, units, expected metadata, repeatable processing, error states and version control. Test missing or inconsistent fields as deliberately as ordinary examples. The system should fail visibly rather than generate a plausible biomarker when its inputs are outside specification. If a human reviewer may alter a contour, the workflow must also specify when that intervention is allowed and retain the original result, user, reason and timestamp. This determines whether a reported trial measurement can be reconstructed and challenged.
Analytical validation asks whether the quantitative output agrees sufficiently with an appropriate reference under the conditions in which it will be used. The first decision is therefore the target measure and comparator.
For segmentation, Dice similarity and related overlap metrics can be informative but incomplete. A minor boundary shift may barely affect a large lesion yet materially change volume or uptake for a small one. If the endpoint is total lesion volume, assessment must include downstream volumetric error and not stop at contour overlap.
Repeatability addresses essentially unchanged conditions; reproducibility addresses changes expected in a multicentre trial. These questions call for measures of bias, variability and agreement, including across scanners or software pipelines where relevant.
Strong correlation does not prove interchangeability. Two methods can move together while reporting systematically different values or different changes from baseline.
A plan should also define unusable scans, low-uptake or very small lesions, outlier handling and acceptable tolerances in relation to the clinical decision. An impressive average performance figure can conceal failure in the cases that matter most.
If an imaging biomarker purports to reflect tumour burden, pharmacodynamic activity or disease response, sponsors must demonstrate why that construct is clinically or biologically meaningful in the population being studied. A result used for exploratory learning needs less evidence than one proposed as a primary efficacy endpoint.
A prognostic association, an association with treatment response and proof of surrogacy for patient benefit are distinct claims. Each needs its own evidence before a model can be called clinically validated.
Disease context, tracer, treatment mechanism, line of therapy and study design all affect interpretation. FDA imaging endpoint process guidance addresses the reliability of acquisition and assessment; it does not convert a technically stable PET number into a recognised surrogate for clinical benefit [6].
A model developed mainly on one institution’s data may encounter different scanner families, acquisition settings and patient characteristics at additional sites. Independent or prospective evaluation should therefore reflect the environments in which the endpoint will actually be used, where practicable. Patient-level separation alone may not be enough to test generalisability if images from the same centres populate training and evaluation datasets.
Before deployment, the sponsor and core lab should agree three operational rules:
• Which input formats, scanner configurations and reconstruction settings are qualified.
• Which software or model version is locked, and how a proposed update will be assessed.
• Who handles failed processing, reader overrides, reanalysis and reconciliation with the clinical dataset.
The test is whether the trial can reconstruct every reported measurement, recognise a material change in the pipeline and prevent silent drift during enrolment.
Hypothetical example – not an Axcellant client case or a validated endpoint. A sponsor proposes AI-assisted whole-body lesion segmentation to calculate a PET-derived total lesion volume as an exploratory biomarker in a multicentre oncology trial.
The first validation mistake would be to report a single pooled segmentation accuracy figure and stop. The sponsor would need to know how the algorithm behaves when there is low tumour uptake, confluent disease, physiological tracer accumulation, metal artefact, different reconstruction kernels and lesions close to the resolution limit. For a longitudinal measure, the relevant question is whether an apparent percentage change exceeds plausible technical variation and whether the algorithm systematically misses new small lesions.
A more useful validation dataset would reserve entire institutions or scanner families for testing, rather than splitting scans from the same patients or centres indiscriminately across training and evaluation datasets. Review would include segmentation and downstream volumetric errors, scan-level failure rate, data outside the trained distribution and subgroup analyses justified by expected failure modes. Performance thresholds should be chosen with input from imaging, clinical and statistical specialists.
If the trial subsequently adopts a software update, the sponsor should assess whether previously processed scans need reanalysis and whether comparability can be preserved. Updating a model because it is “more accurate” on new examples may still change the longitudinal endpoint unless the migration is controlled.
Prof. Piotr J. Słomka, Axcellant Advisory Board
| “AI can make nuclear imaging faster and more consistent, but turning a good demonstration into a measurement a trial can rely on takes careful validation. We need to understand where errors arise, how a method behaves across scanners and patient groups, and whether changes in the output reflect biology rather than processing. AI methods must meet the same standards of validation as any quantitative method developed before them, with particular attention to what the training data comprised and how the method was evaluated externally, on independent data from other sites and scanners. For quantitative PET, that means validating the whole chain, from acquisition to the final quantitative metric.” |
| Failure mode | Why it can bias a trial | A defensible mitigation |
| Patient or site overlap between train and test data | Inflates apparent generalisation | Institution-level separation of testing data |
| Scanner or reconstruction shift | Alters image texture, SUV or lesion boundaries | Qualification and prespecified harmonisation checks |
| Silent processing of out-of-specification images | Produces plausible but unreliable numbers | Automated input validation and visible error states |
| Software drift during enrolment | Changes measurement definition over time | Locked versions, impact assessment and change control |
| Highly imbalanced validation cohorts | Conceals poor performance in difficult cases | Disease- and acquisition-relevant subgroup analyses |
| No downstream biomarker error assessment | Good contours may still yield biased volume/uptake | Evaluate measurement bias and limits of agreement |
| Incomplete audit trails | Prevents reconstruction of the endpoint | Record data, model, reader, override and export versions |
The objective is not to demonstrate that the algorithm never fails. It is to understand the conditions under which it fails, quantify the consequences and ensure that the trial can detect, investigate and manage those failures.
An imaging core lab can coordinate scanner qualification, imaging-charter implementation, data transfer, image QC, analysis pipelines and independent reads as applicable. The sponsor and clinical CRO remain responsible for their own defined duties, including protocol governance, clinical safety oversight, trial data management and vendor oversight. Responsibilities must be established contractually and operationally.
A practical governance model assigns a named owner to the context-of-use statement; one to technical and analytical validation; one to clinical interpretation; and one to controlled software deployment and monitoring. That governance becomes especially important when AI outputs influence real-time eligibility or decisions about continuing treatment. Reprocessing and exception handling should be agreed before first subject in.
Axcellant publicly describes its Imaging Core Lab approach to protocol standardisation, scanner qualification and central workflows [7]. The next step for each sponsor remains a product- and endpoint-specific validation plan.
Wojciech Kula, Founder & CEO, Axcellant
| “Sponsors should not have to choose between innovation and evidentiary discipline. When AI is introduced into a clinical imaging workflow, the business question is whether the resulting information can reliably support the next development decision. That means planning for validation, cross-site performance, controlled software changes and a transparent audit trail from the beginning. Our priority is to connect promising quantitative methods with the level of operational oversight that clinical development requires.” |
A decision-quality dossier should state the context of use; composition of the training data; origin and representativeness of validation datasets, including independent external evaluation; locked algorithm, model and preprocessing versions; comparator method; accuracy, precision and reproducibility findings; subgroup and failure analysis; site and scanner qualification; and rules for human overrides, monitoring, software changes and export. What counts as sufficient depends on the proposed clinical decision. There is no universal AI-accuracy percentage that establishes endpoint validity.
Evaluating AI-assisted PET quantification for clinical development? Discuss context-of-use definition, cross-site reproducibility and imaging endpoint governance with Axcellant’s Imaging Core Lab team.
Reference
[1] US FDA. Drug Development Tool Qualification Programs: Context of Use. View source ↗
[2] Slomka PJ et al. Comparison of clinical tools for measurements of regional stress and rest myocardial blood flow assessed with 13N-ammonia PET/CT. J Nucl Med. 2012;53:171–181. View source ↗
[3] Monroy-Gonzalez AG et al. Software reproducibility of myocardial blood flow and flow reserve quantification in ischemic heart disease: a 13N-ammonia PET study. J Nucl Cardiol. 2020;27:1225–1233. View source ↗
[4] Miller RJH, Slomka PJ. Artificial intelligence in nuclear cardiology: an update and future trends. Semin Nucl Med. 2024;54:648–657. View source ↗
[5] Miller RJH, Slomka PJ. Current status and future directions in artificial intelligence for nuclear cardiology. Expert Rev Cardiovasc Ther. 2024;22:367–378. View source ↗
[6] US FDA. Clinical Trial Imaging Endpoint Process Standards. Final guidance, 2018. View source ↗
[7] Axcellant. Imaging Core Lab in Nuclear Medicine: Standardization, Quality Control, and Regulatory Alignment in Clinical Trials. 2026. View source ↗
A central imaging workflow succeeds when the protocol, imaging charter, site acquisition rules and independent readers make the same assumptions.…