A central imaging workflow succeeds when the protocol, imaging charter, site acquisition rules and independent readers make the same assumptions. This implementation guide focuses on decisions that determine whether central results are traceable and defensible.
A central imaging core lab may receive every scan and still be unable to generate a reliable endpoint. The failure can originate much earlier: an ambiguous imaging charter, differing reader interpretations, incomplete DICOM metadata or a site that reconstructs images outside the qualified parameters. More readers, more software or a second review round will not necessarily repair a design problem.
This article deliberately concentrates on the review and decision workflow rather than revisiting the broader rationale for imaging core labs. Axcellant has already published a separate overview of nuclear medicine standardisation, quality control and regulatory alignment [1]. This guide addresses a narrower, practical question: what must be decided before the first image is submitted for central review?
An imaging charter should translate protocol objectives into executable acquisition, acceptance and interpretation rules. It defines what sites collect, what the core lab checks, how trained readers assess the data and how discrepancies reach resolution. The trial protocol and statistical analysis plan remain controlling documents for clinical objectives and analysis; the charter should implement them consistently; any change to an endpoint definition belongs in those documents.
For PET, SPECT and hybrid imaging, the scope may include acquisition timing, patient preparation, tracer or therapy-related variables, scanner calibration, attenuation correction, reconstruction, transfer format, de-identification, image quality criteria and assessment methodology. Its specificity must match the risk created by imaging in the trial. An exploratory image archive does not require the same reader governance as imaging that determines a pivotal endpoint.
Theranostic programmes add decisions of their own. If a PSMA or somatostatin-receptor PET scan determines eligibility, the charter must state who makes that read, against which criteria and within what turnaround, because a central read that delays dosing affects recruitment. If post-treatment SPECT/CT supports dosimetry, the charter must also fix camera calibration, imaging timepoints relative to administration and the software used to calculate absorbed dose.
| Decision owned before first image | Charter should make explicit |
| Acceptability | What makes a scan technically evaluable or non-evaluable? |
| Timing | What acquisition window is allowed, and what counts as a deviation? |
| Reading | Which rules, lesion conventions and clinical information can readers use? |
| Reader independence | What prevents one reader from seeing another’s assessment? |
| Disagreement | When is adjudication triggered, by whom and with what information? |
| Data lineage | Which image, version, reader and decision produced each variable? |
FDA’s guidance on imaging endpoints emphasises processes that preserve data quality and a verifiable record of acquisition, archiving and interpretation [2]. It does not prescribe one universal blinded independent central review (BICR) design for every study. Sponsors need to justify and document the chosen design.
Consider an illustrative, fictional study in which two central readers reach different response assessments for the same participant. Four very different causes are possible. They may be applying different lesion-selection rules; the underlying reconstruction may differ from the qualified protocol; one reader may have access to a prior scan unavailable to the other; or the images may be technically sound but the lesion boundary genuinely ambiguous.
Treating every discrepancy as a vote to be settled by a third reader hides its source. The workflow should first determine whether there is a data or process defect. Only disagreements that meet the prespecified adjudication definition should proceed to adjudication. A technical-quality issue may require site query, reprocessing or an evaluability decision instead.
An escalation route should distinguish five decisions:
1. Identity and provenance. Confirm patient, timepoint and image set while protecting identifying information.
2. Technical acceptability. Check acquisition and reconstruction compliance before assuming a reader is responsible.
3. Problem classification. Separate missing data and protocol deviations from genuinely ambiguous interpretation.
4. Authorised resolution. Apply only the permitted query, reprocessing, reread or adjudication procedure.
5. Traceability. Preserve source data, versions, decision rationale, who decided and when.
This order matters because a third reader cannot repair a broken data transfer or an uncontrolled reconstruction. Adjudication is reserved for specified interpretative disagreement, once image QC has ruled out a technical cause.
The treating team may need image information promptly to make a participant-care decision, whereas the central review team may be tasked with applying blinded, prespecified endpoint rules. These processes can coexist, but the charter should explain what each reviewer is allowed to see, when an image becomes available and how discrepancies between local and central interpretations are recorded.
When endpoint readers see information that was not meant to be available under the assessment design, the issue is more than an administrative deviation. It may affect the independence of the review and require a documented impact assessment. Conversely, independent endpoint governance must never delay a necessary clinical safety decision.
Quantitative imaging needs a different workflow from qualitative evaluation. A PET-derived uptake or kinetic metric depends on acquisition timing, activity measurement, patient factors, system calibration, reconstruction and software. If longitudinal changes are used, within-patient methodological consistency becomes particularly important.
Professor Piotr J. Słomka and collaborators compared three clinical software tools for myocardial blood-flow quantification using 13N-ammonia PET/CT [3]. Another study, co-authored by Słomka and Professor Juhani Knuuti, investigated software reproducibility of myocardial flow and flow-reserve measures [4]. The first found good agreement between the packages; the second found that, in patients with ischaemia, stress flow and flow reserve differed between packages beyond accepted limits of agreement. Both were cardiac studies, but the lesson applies to any quantitative PET endpoint: agreement between analysis platforms has to be tested in the setting where they will be used.
For a radiopharmaceutical study, the charter should therefore define whether quantitative measurements are produced centrally or locally, how software versions are locked, what constitutes a material processing change and what happens if a site upgrades its scanner during recruitment. Acceptable heterogeneity depends on the intended metric. Phantom-based scanner qualification, such as EARL accreditation or SNMMI Clinical Trials Network scanner validation, is most useful when it tests the specific measurement risk of the intended metric.
Prof. Piotr J. Słomka, Axcellant Advisory Board
| “A reproducible imaging endpoint starts long before two readers agree. It depends on the acquisition protocol, image reconstruction, quantitative software and interpretation rules producing sufficiently comparable results across the study. A good imaging charter makes these dependencies visible and testable, locks software versions and treats any mid-study software or scanner upgrade as a change to the measurement. When data are inconsistent, the first question should be where the measurement pathway changed, and only then which reader should decide.” |
Data lineage begins before the first participant is scanned. The sponsor and core lab need an agreed dataset specification, visit reconciliation rules and a method to detect incomplete image series. A pilot transfer should test de-identification, DICOM metadata, acquisition parameters and the assignment of each study to the right participant and timepoint.
Reader training then tests whether the written criteria work on “real world” cases. If reviewers interpret a rule differently, the charter should be clarified under version control before the trial depends on it.
During conduct, operational quality indicators are more useful than a simple count of scans received. Rejection reasons, delayed transfers, recurring site deviations, unresolved image queries and unusual adjudication patterns can reveal a systematic problem while training or corrective action remains possible. These signals should be investigated without changing endpoint interpretation opportunistically.
At data lock, each endpoint variable should be traceable to an accepted DICOM study, the applicable charter and software versions, initial reads, any adjudication and the exported dataset. The goal is an audit trail from which a reviewer can reconstruct how each measurement was produced.
Practical test: Select one disputed endpoint value and follow it backwards to the original image. If the explanation depends on an undocumented email or an individual’s memory, governance is incomplete.
Wojciech Kula, Founder & CEO, Axcellant
| “Sponsors should not discover during database lock that the history of an imaging decision cannot be reconstructed. Our view is that independent image review must be designed as a controlled operational process from the outset, with clear roles for sites, reviewers, data management and the clinical team. The commercial value of an imaging core lab lies in reliable, reviewable evidence–not only in the volume of scans it can process.” |
A full rehearsal can expose missing decisions more efficiently than another written sign-off. Use a representative test dataset and follow it from the proposed site through final endpoint export:
• Transfer and reconcile the scan, including de-identification and required metadata.
• Apply the stated technical QC and document any rejection or site query.
• Route the accepted image to independent readers under the agreed blinding rules.
• Trigger a planned discrepancy, execute the authorised resolution and check the audit trail.
• Export the imaging variable and reconcile it against the receiving trial dataset.
Only after that end-to-end test should the team regard the charter as operationally executable.
Need an imaging charter or an independent central image-review process for a radiopharmaceutical study? Discuss a protocol-specific model with Axcellant’s Imaging Core Lab team.
Reference
[1] Axcellant. Imaging Core Lab in Nuclear Medicine: Standardization, Quality Control, and Regulatory Alignment in Clinical Trials, February 2026. View source ↗
[2] US FDA. Clinical Trial Imaging Endpoint Process Standards, final guidance, April 2018. View source ↗
[3] Slomka PJ et al. Comparison of clinical tools for measurements of regional stress and rest myocardial blood flow assessed with 13N-ammonia PET/CT. J Nucl Med. 2012;53:171–181. View source ↗
[4] Monroy-Gonzalez AG et al. Software reproducibility of myocardial blood flow and flow reserve quantification in ischemic heart disease: a 13N-ammonia PET study. J Nucl Cardiol. 2020;27:1225–1233. View source ↗